# Philips 867041

> Source: https://shop.ezmedsource.com/products/philips-867041

## At a glance

- **Manufacturer:** Philips
- **Model:** 867041
- **Category:** Patient Monitor
- **Modality:** Module

## Description

The Philips 867041 is a CO₂ extension module for Philips IntelliVue patient monitors. It connects via the Multi-Measurement Server (MMS) and measures end-tidal CO₂ (EtCO₂), inspired CO₂ (FiCO₂), and respiratory rate (RR) using Microstream capnography technology. The module supports single- and dual-port sampling lines, enabling simultaneous capnography and oxygen delivery on a single patient line. It reaches full measurement accuracy in under 30 seconds and works with adult, pediatric, and neonatal patients. Alarm thresholds for CO₂ and respiratory rate are customizable. Compatible with IntelliVue X2, MP, and MX series monitors.

- Microstream low-flow CO₂ measurement technology
- Dual-port sampling line support
- Warm-up time under 30 seconds
- Customizable CO₂ and RR alarms
- Supports all patient populations (adult, pediatric, neonatal)

## Tech Info

- **Model Number:** 867041 (Module Code: M3015A or M3015B)
- **Measurement Technology:** Microstream capnography
- **Parameters Measured:** End-tidal CO₂ (EtCO₂), inspired CO₂ (FiCO₂), respiratory rate (RR)
- **Sample Line Compatibility:** Single- and dual-port Microstream sampling lines
- **Warm-Up Time:** Less than 30 seconds to full accuracy
- **Alarm Thresholds:** Customizable for EtCO₂ and RR
- **Monitor Compatibility:** Philips IntelliVue MMS, X2, MP, and MX series
- **Design:** Compact module for direct attachment to MMS; lightweight, designed for mobile and bedside use

The module operates in continuous monitoring mode during procedures, transport, and general care.

## Common Problems

- No communication with monitor after connection or power-up.
- CO₂ measurement absent or incorrect; verify sample line is not occluded.
- Parameter display blanked out; check parameter connector for damage or loose seating.
- CO₂ equipment malfunction; replace sampling line or main module if measurement fails.
- Parameter connector damaged; inspect for bent pins or physical wear.
- Software incompatibility with monitor firmware version.

## Specifications

| Spec | Value |
| --- | --- |
| features | Cardiac Output B05,Hardware D,Invasive Pressure Temp B06 |
| Parameters | EtCO₂, FiCO₂, RR |
| Model Number | 867041 (M3015A or M3015B) |
| Warm-Up Time | Less than 30 seconds |
| Monitor Compatibility | Philips IntelliVue MMS, X2, MP, MX series |
| Measurement Technology | Microstream capnography |
| Sample Line Compatibility | Single- and dual-port Microstream lines |

## Vendor offers

### Pioneer Biomedical

- **TradeIn** · Quote on request · availability: AvailableOnRequest
- **Repair** · Quote on request · availability: AvailableOnRequest
- **Rent** · Quote on request · availability: AvailableOnRequest
- **Purchase** · Quote on request · availability: InStock

### Prescott's Med

- **Repair** · Quote on request · availability: AvailableOnRequest
- **Purchase** · Quote on request · availability: InStock

## Documents