# Philips IntelliVue G7 (866173)

> Source: https://shop.ezmedsource.com/products/philips-intellivue-g7-866173

## At a glance

- **Manufacturer:** Philips
- **Model:** IntelliVue G7
- **Model number:** 866173
- **Category:** Patient Monitor
- **Modality:** Module

## Description

The Philips IntelliVue G7 Anesthesia Gas Module (866173) is a dual-slot parameter module for IntelliVue patient monitors that measures anesthetic gases and respiratory parameters. It samples anesthetic agents (sevoflurane, isoflurane, halothane, desflurane, enflurane), nitrous oxide, carbon dioxide, oxygen, and airway respiratory rate from the patient breathing circuit. The module features a reusable, rotatable water trap that can be positioned flexibly on integrated or rack-mounted monitor setups. It requires IntelliVue monitors (MP40, MP50, MP80, MP90, MX500–MX800 series) running software level L.0 or later.

## Tech Info

- **Sample Flow Rate:** 200 ml/min ±20 ml/min
- **Sevoflurane Accuracy:** ±0.5% (2.00% expected)
- **Isoflurane Accuracy:** Not specified
- **Halothane Accuracy:** Not specified
- **Desflurane Accuracy:** ±0.5% (2.00% expected)
- **Enflurane Accuracy:** Not specified
- **N₂O Accuracy:** ±6.2% (43% expected)
- **CO₂ Accuracy:** ±0.7% (5.00% expected)
- **O₂ Accuracy:** ±3% (50% expected)
- **Water Trap:** Reusable, rotatable design
- **Minimum Software Requirement:** IntelliVue L.0 or later

Refer to the Philips G7 Checkout Procedure for detailed leak test and calibration instructions.

## Common Problems

- Gas sampling line occlusion or blockage.
- Low sample flow rate.
- Pump motor does not activate.
- Module remains in warmup mode and does not transition to measurement mode.

## Specifications

| Spec | Value |
| --- | --- |
| Model | 866173 |
| Manufacturer | Philips |
| Product Line | IntelliVue G7 Anesthesia Gas Module |
| Minimum Software | L.0 |
| Sample Flow Rate | 200 ml/min ±20 ml/min |
| Compatible Monitors | IntelliVue MP40, MP50, MP80, MP90, MX500–MX800 |

## Vendor offers

### Pioneer Biomedical

- **TradeIn** · Quote on request · availability: AvailableOnRequest
- **Repair** · Quote on request · availability: AvailableOnRequest
- **Rent** · Quote on request · availability: AvailableOnRequest
- **Purchase** · Quote on request · availability: InStock

### Prescott's Med

- **Repair** · Quote on request · availability: AvailableOnRequest
- **Purchase** · Quote on request · availability: InStock

## Documents

- [Philips_G7_Checkout Procedure.pdf](https://storage.googleapis.com/ezmedsource-product-images/product-manuals/121f0a2f710ebb21.pdf) — manual

---