# Philips M1019A, M1013A - G5 Gas Module (M1019A, M1013A)

> Source: https://shop.ezmedsource.com/products/philips-m1019a-m1013a-g5-gas-module-m1019a-m1013a

## At a glance

- **Manufacturer:** Philips
- **Model:** M1019A, M1013A - G5 Gas Module
- **Model number:** M1019A, M1013A
- **Category:** Anesthesia Machine

## Description

The Philips M1013A and M1019A are anesthetic gas modules for the MP and MX series patient monitors. They measure end-tidal concentrations of CO2, O2, N2O, and anesthetic agent gases via a pneumatic sampling line (nominal flow 200 ml/min). The module requires monitor software version D.0 or higher. The M1013A/M1019A is a smaller form factor compared to the earlier M1026B gas module. Annual preventive maintenance and checkout procedures are recommended per Philips guidelines.

## Tech Info

- **Gas Parameters:** CO2, O2, N2O, anesthetic agent (desflurane).
- **Sampling Flow:** 200 ml/min ± 20.
- **Expected Accuracy (End-Tidal):** Desflurane 2.00% ± 0.5%, CO2 5.00% ± 0.7%, N2O 43% ± 6.2%, O2 50% ± 3%.
- **Software Requirement:** Monitor software level D.0 or higher.
- **Compatible Monitors:** MP series, MX series Philips patient monitors.

Leak testing and checkout procedures are documented in the service manual.

## Common Problems

- Low or unstable CO2 readings.
- Occlusion in the sampling line or gas circuit.
- Pneumatic pump fails to activate or run.
- Module remains in warmup mode and does not transition to measurement.
- O2 readings lower than expected or inconsistent.

## Specifications

| Spec | Value |
| --- | --- |
| Model | M1013A, M1019A |
| O2 Accuracy | 50% ± 3% |
| CO2 Accuracy | 5.00% ± 0.7% |
| N2O Accuracy | 43% ± 6.2% |
| Gas Parameters | CO2, O2, N2O, desflurane |
| Sampling Flow Rate | 200 ml/min ± 20 |
| Desflurane Accuracy | 2.00% ± 0.5% |
| Minimum Software Version | D.0 |
| Compatible Monitor Series | MP, MX |

## Vendor offers

### Pioneer Biomedical

- **TradeIn** · Quote on request · availability: AvailableOnRequest
- **Repair** · Quote on request · availability: AvailableOnRequest
- **Rent** · Quote on request · availability: AvailableOnRequest
- **Purchase** · Quote on request · availability: InStock

### Prescott's Med

- **Repair** · Quote on request · availability: AvailableOnRequest
- **Purchase** · Quote on request · availability: InStock

## Documents

- [Philips_M1013 M1019A_CheckOutProcedure.pdf](https://storage.googleapis.com/ezmedsource-product-images/product-manuals/23739546e6decd86.pdf) — manual

---
Generated from https://shop.ezmedsource.com/products/philips-m1019a-m1013a-g5-gas-module-m1019a-m1013a. RFQ flow on EzMedSource is the primary commerce primitive — submit a quote request to receive responses from every vendor matching your line items.